![Panametrics DigitalFlow XGF868i Service Manual Download Page 50](http://html1.mh-extra.com/html/panametrics/digitalflow-xgf868i/digitalflow-xgf868i_service-manual_3407914050.webp)
Appendix B. Calibration and Testing with PanaView
42
DigitalFlow™ XGF868i Service Manual
B.4.1 Analog Outputs
Additional analog outputs may be added to the XGF868i by installing an appropriate option card in Slot 1. This option
card contains two or four analog outputs, which are designated as A, B, C and D. Both the zero-point and full-scale
values for each output must be calibrated. After calibrating the outputs, which have a resolution of 5.0
A (0.03% full
scale), their linearity should be tested.
Note:
The zero point of the analog output may be set for either 0 mA or 4 mA. However, the calibration procedure
always uses the 4 mA point, as the meter will extrapolate this value to obtain the 0 mA point.
Prepare for calibration by inserting an ammeter in series with the load on Output A of the option card. See
to identify the OUT(+) and RTN(-) pins for Output A on terminal block J2.
2. Double-click on either
Output A
or
Output B
to open the
Analog Output
menu. (Output A will be used as an exam
-
ple in this manual.)
IMPORTANT:
The procedure for calibrating
Output B
is identical to that for calibrating
Output A
. However, be sure to
reconnect the ammeter on terminal block
J2
.
for the correct pin numbers.
3. Double-click on
4 milliamps
to calibrate the low end of the output range.
4. Double-click on
4 mA UP
or
4 mA DOWN t
o adjust the ammeter reading UP or DOWN, until a 4 mA reading is
achieved, or double-click on
4 mA Numeric Cal,
and directly enter the mA reading in the right pane.
5. Once you have achieved a 4 mA reading, double-click on either
4 mA STORE
to store the calibration, or on
4 mA
ABORT
to exit the menu without storing the calibration.
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 4 or 20 mA setting, contact the factory for
assistance.
6. PanaView returns to the
Analog Output
window. Double-click on
20 milliamps
, and repeat steps 3 through 5 to
calibrate the high end of the output range.
Note:
Skip the following prompt, if linearity testing of the output will not be performed at this time.
7. Double-click on
% Test
to test the linearity of the currently selected analog output.
8. Check the ammeter reading at
50% output
. Then, enter a different
output
(0-100%) and double-click on
percent
.
Check the ammeter reading at this setting and press
[Exit Page]
when done.
lists the expected ammeter readings at various
% Full Scale
settings, for both 4-20 mA and
0-20 mA scales. Refer to this table to verify the accuracy of the ammeter readings taken above. If the linearity test
readings are not within 5
A of the values listed in the table, check the accuracy and wiring of the ammeter. Then,
repeat the low and high end calibrations. If the analog output still does not pass the linearity test, contact the factory
for assistance.
Click
[Exit Page]
to return to the
Slot 1
window and calibrate additional inputs/outputs, or click
[Exit Page]
again to close
the
Calibration/Test
menu.
Summary of Contents for DigitalFlow XGF868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Page 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 71: ......