![Panametrics DigitalFlow XGF868i Service Manual Download Page 32](http://html1.mh-extra.com/html/panametrics/digitalflow-xgf868i/digitalflow-xgf868i_service-manual_3407914032.webp)
Chapter 4. Parts Replacement
24
DigitalFlow™ XGF868i Service Manual
4.4 Replacing the Fuse
IMPORTANT:
This procedure applies only to the fuse on the DC power supply. The fuse on the AC power supply is not
field-replaceable; please contact Panametrics for replacement.
If the fuse in the XGF868i has been blown, complete the following steps to install a new fuse:
1.
Complete the steps in
“Removing the Circuit Board Assembly” on page 22
.
2.
Place the circuit board assembly on its side, with the terminal board on the left and the interconnect board on
the right. Rotate the assembly until the power supply board is facing upward.
3.
Locate the fuse along the left side of the power supply board, just below the terminal board (refer to
Illustration 6
in
).
4.
Pull the plastic cover off the fuse holder and remove the old fuse.
5.
Obtain a new fuse of the same rating and type. Use only a type 3AG (1-1/4” x 1/4”) Slo-Blo fuse, having a rating as
indicated in
below.
6.
Press the new fuse into the fuse holder and reinstall the plastic fuse cover.
To complete the fuse replacement, proceed to
“Installing the Circuit Board Assembly” on page 28
Note:
Be sure to record the fuse replacement in Appendix A,
Service Record
.
Table 4: Line Voltages & Fuse Ratings
Line Voltage
Fuse Rating
12-28 VDC
2.0 A, Slo-Blo
85-250 VAC
Not field replaceable
(contact Panamtrics)
Summary of Contents for DigitalFlow XGF868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Page 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 71: ......