DigitalFlow™ XGF868i Service Manual
55
Appendix B. Calibration and Testing with PanaView
Table 10: Available Measurement Parameters
Option Bar
Description
Good
Bad
VEL
Displays the flow velocity
N.A.
N.A.
VOLUM
Displays the volumetric flow
N.A.
N.A.
+TOTL
Displays the forward totalized volume flow
N.A.
N.A.
-TOTL
Displays the reverse totalized volume flow
N.A.
N.A.
TIME
Displays the total flow measurement time
N.A.
N.A.
MDOT
Displays the mass flow
N.A.
N.A.
+MASS
Displays the forward totalized mass flow
N.A.
N.A.
-MASS
Displays the reverse totalized mass flow
N.A.
N.A.
SS up
Displays the signal strength for the upstream transducer
50–75
<50 or >75
SS do
Displays the signal strength for the downstream transducer
50–75
<50 or >75
SNDSP
Displays the measured speed of sound in the gas
N.A.
N.A.
Tup
Displays the upstream ultrasonic signal transit time
N.A.
N.A.
Tdown
Displays the downstream ultrasonic signal transit time
N.A.
N.A.
DELTA
Displays the transit time difference between the upstream and
downstream signals
N.A.
N.A.
Tot K
Displays the total K factor.
N.A.
N.A.
PEAK%
Displays the percentage of peak (set to +50 by default)
N.A.
N.A.
Qup
Displays the signal quality for the upstream transducer
>1200
–400 to +400
Qdown
Displays the signal quality for the downstream transducer
>1200
–400 to +400
AMPup
Displays the signal amplitude value of the upstream transducer
24 ± 5
<19 or >29
AMPdn
Displays the signal amplitude value of the downstream transducer
24 ± 5
<19 or >29
CNTup
Displays the AGC DAC count for the upstream gain setting.
N.A.
N.A.
CNTdn
Displays the AGC DAC count for downstream gain setting.
N.A.
N.A.
P#up
Displays signal peaks for the upstream transducer.
100-2300 <100 or >2300
P#dn
Displays signal peaks for the downstream transducer.
100-2300 <100 or >2300
TEMP
Displays the gas temperature (from 0/4-20 mA input).
N.A.
N.A.
PRESR
Displays the gas pressure (from 0/4-20 mA input).
N.A.
N.A.
Mw
Displays the molecular weight.
N.A.
N.A.
z
Displays the compressibility.
N.A.
N.A.
AcVOL
Displays actual volumetric flow.
N.A.
N.A.
StVOL
Displays standard volumetric flow.
N.A.
N.A.
Tu S
1
Displays Skan transit time upstream.
N.A.
N.A.
Td S
1
Displays Skan transit time downstream.
N.A.
N.A.
DT S
1
Displays Skan Delta T.
N.A.
N.A.
Tu M
1
Displays Measure transit time upstream.
N.A.
N.A.
Td M
1
Displays Measure transit time downstream.
N.A.
N.A.
DT M
1
Displays Measure Delta T.
N.A.
N.A.
Vinst
Displays the instantaneous velocity.
N.A.
N.A.
Summary of Contents for DigitalFlow XGF868i
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 24: ...Chapter 2 Error Codes 16 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 40: ...Chapter 4 Parts Replacement DigitalFlow XGF868i Service Manual 32...
Page 70: ...Warranty 62 DigitalFlow XGF868i Service Manual no content intended for this page...
Page 71: ......