DigitalFlow™ GS868 Programming Manual (2-Channel)
77
Chapter 6. Serial Communications
6.5.3 Multi-Point Wiring
The standard point-to-point wiring configuration for the serial interface converter may be modified to permit the use
of a multi-point wiring arrangement. In a multi-point RS485 system, one flowmeter (the master) is connected to the
personal computer, while a number of additional flowmeters (the slaves) are chained together and connected to the
master flowmeter. In order to implement such a system, the
DIP
switch settings inside each serial interface converter
must be changed.
IMPORTANT:
The serial interface converter in the last slave unit in the chain should not be reconfigured.
6.5.3.1
Reconfiguring a Serial Interface Converter
To reconfigure a serial interface converter for multi-point wiring, complete the following steps:
1.
Disconnect the
main power to the electronics console and open the cover.
2.
Remove the clear plastic
shroud that covers the electrical connectors.
3.
Remove the serial interface converter
mounting bracket by removing the standoff located just below the
RS232
terminal block and the grounding screw to its left (see Figure 9 on page 75).
4.
Loosen the two screws that fasten the
DB9
connector to the mounting bracket, and remove the
serial interface
converter from the bracket.
5.
Use a small screwdriver to pry open the
plastic case of the serial interface converter, as shown in Figure 10 on
The serial interface converter contains a main printed circuit board and a daughter printed circuit board. The
daughter board has a small terminal block attached to it and the main board has a
DIP
switch (
SW1
) mounted near
the center of the board.
6.
Locate the switch assembly on the main board, and move the switch in position 1 from ON to OFF. The standard
ON state for this switch is for point-to-point operation, while the OFF state is needed for multi-point operation.
See Table 28 on page 78 for the proper settings of all four switches on the switch assembly.
WARNING!
Dangerous voltages exist within the electronics console. Do not perform any wiring
operations until the main power to the unit has been disconnected.
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......