Chapter 1. Programming Site Data
20
DigitalFlow™ GS868 Programming Manual (2-Channel)
2.
At the
Static Density?
prompt, press
[F1]
=
NO
or
[F2]
=
YES
. If you press
NO
, the GS868 returns to the
ADVAN
window.
3.
Enter the
Fluid Density of the gas to be measured and press
[ENT]
. (The meter will only accept values from
0.00001-100 lb/ft
3
.)
The GS868 program returns to the
ADVAN
window.
1.7.6.2
Entering Multiple K Factors
In the
MULTK
option, you can enter velocity or Reynolds Correction values into a K-Factor table. You can thus “curve
fit” velocity or Reynolds calibration multiple data points (from several different data sources or flow variables) to the
flow rate reported by the GS868.
1.
Press
[F2]
to enter the
Multiple K Factors option.
2.
Press
[F1] = NO
to disable or
[F2] = YES
to enable
Multiple K-factors.
• If you press
NO
, the GS868 returns to the
ADVAN
window.
3.
If you press
YES
, the program asks for the
Custom type of data. Press
[F1]
to select velocity data, or
[F2]
to select
Reynolds data.
a.
If you select a Reynolds table, the meter asks for the
Kinematic Viscosity. Enter the appropriate value, and press
[ENT]
.
4.
Press
[F1] = NO
to retain the current K-factor table and return to the
ADVAN
window, or press
[F2] = YES
to
Edit
The K-factor Table.
Note:
If velocity vs. K-factor data was not provided with the Model GS868, the K-factor table can not be edited.
a.
Enter the
Number of K-factors to be entered into the table. Press
[ENT]
. (The meter will only accept values from 2 to
20.)
IMPORTANT:
When editing the K-factor table, the velocities must be entered in increasing order.
b.
Enter the
Velocity (or Reynolds) Value for K-factor number “x”. Press
[ENT]
. (The meter will only accept values
from –30,000 to +30,000 ft/sec.)
c.
Enter the
K-factor corresponding to velocity or Reynolds number “x”. Press
[ENT]
.
d.
Repeat steps b and c until all of the data points have been entered.
Procedure Options
After completing the above steps, the meter returns to the
SETUP
prompt. Continue as follows:
• To continue programming the meter, refer to the menu maps in Appendix A and navigate to the desired menu.
Then, proceed to the appropriate section of this manual for instructions.
• To leave the
User Program and retain the previous settings, press
[EXIT]
twice and then press
[F1] = NO
at the
SAVE prompt. Any programming changes will be discarded and you will be returned to the data display.
• To leave the
User Program and return to measurement mode, press
[EXIT]
twice and then press
[F2] = YES
at
the
SAVE prompt. Your programming changes will be entered into the meter’s memory, and you will be
returned to the data display.
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......