DigitalFlow™ GS868 Programming Manual (2-Channel)
15
Chapter 1. Programming Site Data
1.7.4.3
Entering the Temperature Input
5.
At the
Temperature Input prompt, press
[F1]
to enter a constant temperature value or press
[Fx] to select the
option card in Slot x that will supply the live temperature input.
Note:
Each slot that contains an option card with an analog input assigned to
TEMP
or an
RTD
input will appear on
the option bar. If the process temperature is stable, a fixed value may be used, but most applications require
a live temperature input.
6.
Do one of the following:
• If you selected
FIXED
, enter the known fixed process temperature. Press
[ENT]
. (The meter will only accept
values from –148° to 662°F.)
• If you selected
SLOT X,
a.
Press
[F1]
to select input A or press
[F2]
to select input B.The inputs were labeled during setup.
Note:
The set up of input A is used as an example. Identical procedures would be used to set up input B.
b.
Enter the
Base Temperature (from –148° to 662°F), and press
[ENT]
. The ratio of this value to the actual
temperature is used to calculate the standard volumetric flow.
1.7.4.4
Entering the Pressure Input
7.
At the
Pressure Input prompt, press
[F1]
to enter a constant pressure value or press
[Fx]
to select the option card
in Slot X that will supply the live pressure input.
Note:
Each slot that contains an option card with an analog input assigned to
PRESR
will appear on the option bar.
If the process pressure is stable, a fixed value may be used, but most applications require a live pressure
input.
8.
Do one of the following:
• If you selected
FIXED
, enter the known fixed process pressure. Press
[ENT]
. (The meter will only accept values
from 0–5000 psia.)
• If you selected
SLOT X,
a.
Press
[F1]
to select input A or press
[F2]
to select input B.The inputs were labeled during setup.
b.
Enter the
Base Pressure (standard) value for the process, and press the
[ENT]
key. (The meter will only accept
values from 0–5000 psia.)
c.
Press
[F2] = YES
to activate the
Low Pressure Switch, or press
[F1] = NO
to turn the switch off.
• If you selected
NO
, proceed to step 9.
• If you selected
YES
, enter the
Pressure Limit (the setpoint for the low pressure switch) and press
[ENT]
. The
acceptable range is 0 to 5,000 psia. If the pressure drops below the pressure limit, the GS868 will stop taking
readings.
1.7.4.5
Entering the Quality Input Value
9.
At the
Quality Input prompt, press
[F1]
to enter a constant steam quality value or press
[Fx]
to select the option
card in Slot X that will supply the live steam quality input. This value is used for mass flow calculations when the
temperature and pressure inputs indicate that the steam is saturated. Unless a different value is accurately
known, accept the default value.
Note:
Each slot that contains an option card with an analog input that has been programmed as
SPEC
in the
OPTN
submenu appears on the option bar. Normally, a fixed value of 1.0 (100% steam) should be used for the steam
quality value.
10.
Do one of the following:
• If you selected
FIXED
, enter the known fixed steam quality. Press
[ENT]
. (The meter will only accept values from
0.0000 to 1.0000.)
• If you selected
SLOT X
,
a.
Press
[F1]
to select input A or press
[F2]
to select input B.The inputs were labeled during setup.
Procedure Options
After completing the above steps, the meter returns to the main I/O prompt. Continue as follows:
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......