DigitalFlow™ GS868 Programming Manual (2-Channel)
27
Chapter 1. Programming Site Data
1.8.3.2
Setting Up Input, Output and Option Cards
The Model GS868 has two built-in analog outputs, which are assigned to
Slot 0
. Also, a variety of option cards may be
installed in the six expansion slots. See Chapter 1,
Installation, of the Startup Guide for a complete description of the
available option cards.
Note:
For this discussion,
Slot x
will be used to reference any of the specific expansion slots
(Slot 1 - Slot 6)
. If the
number of a slot with an installed option card does not appear, the card may not have been initialized or it
may be defective. Call the factory for assistance.
• Use the
]
and
] and the
[F1]-[F4]
keys to select the desired slot number. (Only those slots which contain
an option card will appear on the option bar.)
Refer to the menu map in Figure 14 on page 86
, and complete the steps on the pages referenced below to setup
and/or scale the Model GS868’s analog inputs and outputs.
• Analog Outputs — page 27
• Alarms — page 27
• Totalizer/Frequency Outputs —page 28
• Analog Inputs — page 28
• RTD Inputs — page 29
Setting Up Analog Outputs
1.
At the
Slot X Outputs prompt, press
[F1]
to set up output A or
[F2]
to set up output B.
Note:
The set up of output A is used as an example. Identical procedures would be used to set up output B.
2.
Press
[F1] = OFF
to disable output A and return to the previous prompt, or press
[F2] = 0-20 m or [F3] = 4-20 m
to
specify the desired range for output A.
3.
Use the
]
and
] and
[F1]-[F4]
keys to select a
Channel option for the data source.
4.
Use the
]
and
] and
[F1]-[F4]
keys to specify the desired
Output Measurement parameter.
5.
Enter the
Zero value for the low end of the chosen output range. Press
[ENT]
.
6.
Enter the
Full value for the high end of the chosen output range. Press
[ENT]
.
7.
Press
[F1]
-
[F2]
to set up another output or press
[EXIT]
to select another slot for set up.
Setting up Alarms
1.
Use the
[F1]-[F3]
keys to set up alarm relays A, B, or C, respectively.
Note:
The set up of alarm A is used as an example. Identical procedures would be used to set up the other alarms.
2.
Press
[F1] = OFF
to disable Alarm A and return to the previous prompt, or press
[F2] = HIGH, [F3] = LOW or [F4]
=
FAULT
to specify the type for Alarm A.
3.
Press
[F1] = NO
for standard operation or
[F2] = YES
for
Failsafe operation.
4.
Use the
]
and
] and
[F1]-[F4]
keys to select a
Channel option for the data source.
5.
Do one of the following:
• If you selected
FAULT
,
proceed to step 6.
• If you selected
HIGH
or
LOW
,
complete the steps below.
a.
Use the
] and
] and
[F1]-[F4]
keys to specify the desired
Output Measurement parameter.
b.
Enter a value for the trigger point of the alarm and press
[ENT]
.
6.
To select the
Fault Type that will trigger the alarm, press
[F1] = FLOW, [F2] = OTHER
(non-flow), or
[F3]= ALL
.
7.
Press
[F1]
-
[F3]
to set up another alarm or press
[EXIT]
to select another slot for set up.
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Programming Manual 2 Channel...
Page 91: ...DigitalFlow GS868 Programming Manual 2 Channel 81 Appendix A Menu Maps...
Page 92: ...82 DigitalFlow GS868 Programming Manual 2 Channel...
Page 100: ...DigitalFlow GS868 Programming Manual 2 Channel 90...
Page 104: ...94 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 122: ...112 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 136: ...126 DigitalFlow GS868 Programming Manual 2 Channel no content intended for this page...
Page 140: ...Index 4 DigitalFlow GS868 Programming Manual 2 Channel...
Page 144: ...Chapter 2 DigitalFlow GS868 Programming Manual 2 Channel...
Page 145: ......