Appendix B. Optional Enclosures
44
DigitalFlow™ GM868 Service Manual
B.3 Rack Mount Fuse Replacement
If it has been determined that the fuse in the Model GM868 requires replacement, refer to Figure 10 on page 51 and
Figure 11 on page 52, and complete the following steps:
1.
On the rear panel of the meter, set the power switch to the OFF position and pull the power cord out of its recep
-
tacle.
2. Locate the black plastic fuse cover on the back panel at the bottom of the power cord receptacle and pull it
straight out.
3. Replace the defective fuse with a new one of the same rating and type. Use only 5 x 20 mm Slo-Blo fuses, having
a rating as indicated in Table 7 below and on the rear panel label.
4. Reinstall the black plastic fuse cover and insert the power cord back into the receptacle. Then, position the power
switch in the ON position.
The Model GM868 flowmeter may now be placed back into service. Reconnect the main power and resume taking
measurements.
Note:
Be sure to record the fuse replacement in Appendix A, Service Record.
WARNING! The main power to the Model GM868 must be disconnected before proceeding.
Table 7: Line Voltages & Fuse Ratings
Line Voltage
Fuse Rating
100-120 VAC
1.0 A, Slo-Blo
220-240 VAC
0.5 A, Slo-Blo
12-28 VDC
3.0 A, Slo-Blo
Summary of Contents for DigitalFlow GM868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 9: ...DigitalFlow GM868 Service Manual vii Contents...
Page 10: ...Contents viii DigitalFlow GM868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Page 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Page 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Page 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Page 71: ......