DigitalFlow™ GM868 Service Manual
7
Chapter 1. Calibration
1.5 Calibrating the Analog Inputs (cont.)
For this discussion, assume that the option card has been installed in
Slot x
.
Note:
The zero point of the analog input may be set for either 0 mA or 4 mA. However, the calibration procedure
always uses the 4 mA point, as the meter will extrapolate this value to obtain the 0 mA point.
1.5.1
Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate
the desired slot
. (The option bar will include a slot listing for each installed option card.)
3.
Press
[Fx]
to select the desired input.
The procedure for calibrating
Input A
is identical to that for calibrating
Input B
. However, when calibrating
Input B
, be
sure a calibrated current source is connected to the appropriate pins on the terminal block. See Figure 3 on the
previous page for the correct pin numbers. To calibrate a 0/4-20 mA input card, see the section below.
Note:
To calibrate an RTD input card, refer to page 9.
4.
Press
[F1]-[F4]
to select the reference point to be calibrated.
Do one of the following:
• If you selected
[F1] (aLOW)
, proceed to
aLOW Option
on the next page.
• If you selected
[F2] (aHIGH)
, proceed to
aHIGH Option
on the next page.
• If you selected
[F3] (4 mA)
, proceed to
4 mA Option
on the next page.
• If you selected
[F4] (20 mA)
, proceed to
20 mA Option
on the next page.
Summary of Contents for DigitalFlow GM868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 9: ...DigitalFlow GM868 Service Manual vii Contents...
Page 10: ...Contents viii DigitalFlow GM868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Page 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Page 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Page 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Page 71: ......