Chapter 3. Diagnostics
24
DigitalFlow™ GM868 Service Manual
3.2 Displaying Diagnostic Parameters (cont.)
Table 3 below lists the available diagnostic parameters for the Model GM868. The first column in the table shows the
parameter as it appears on the option bar, while the second column shows the parameter as it appears in the
prompt area after it has been selected.
Ch1 label gSITE
Press the
]
and
keys and the appropriate
[Fx]
key to display a
different diagnostic parameter or to
EXIT
the
Diagnostics Menu.
Ch1 DN Transit usec
519.7
(error codes appear here)
Tdown
DELTA
PEAK
%
Qup
Table 3: Available Diagnostic Parameters
Option Bar
Screen Display
Description
Good
Bad
SS up
UP Sig Strength
Displays the signal strength for the
upstream transducer.
50–75
<50 or >75
SS do
DN Sig Strength
Displays the signal strength for the
downstream transducer.
50–75
<50 or >75
SNDSP
Soundspeed Ft/s Displays the measured sound speed of
the fluid.
N.A.
N.A.
Tup
UP Transit S usec Displays the upstream transit time of
the ultrasonic signal in microseconds.
N.A.
N.A.
Tdown
DN Transit usec
Displays the downstream transit time of
the ultrasonic signal in microseconds.
N.A.
N.A.
DELTA
DeltaT usec
Displays the difference in µsec between
upstream and downstream transit
times.
N.A.
>10,000
nsec
PEAK%
PEAK%
Displays the percentage of peak
(set to +50 by default).
N.A.
N.A.
Qup
UP Signal Q
Displays the signal quality for the
upstream transducer.
1200
–400 to
+400
Qdown
DN Signal Q
Displays the signal quality for the
downstream transducer.
1200
–400 to
+400
AMPup
UP Amp Discrim
Displays the value for the amplitude
discriminator of the upstream
transducer.
24 ± 5
<19 or >29
AMPdn
DN Amp Discrim Displays the value for the amplitude
discriminator of downstream
transducer.
24 ± 5
<19 or >29
CNTup
UP DAC Counts
Displays the AGC DAC count for
upstream gain setting.
N.A.
N.A.
CNTdn
DN DAC Counts
Displays the AGC DAC count for
downstream gain setting.
N.A.
N.A.
P#up
UP +- Peak
Displays signal peaks for the upstream
transducer.
100-230
0
<100 or
>2300
P#dn
DN +- Peak
Displays signal peaks for the
downstream transducer.
100-230
0
<100 or
>2300
Summary of Contents for DigitalFlow GM868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 9: ...DigitalFlow GM868 Service Manual vii Contents...
Page 10: ...Contents viii DigitalFlow GM868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Page 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Page 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Page 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Page 71: ......