Chapter 1. Calibration
2
DigitalFlow™ GM868 Service Manual
1.3
Calibrating the Analog Outputs
Every Model GM868 flowmeter includes two built-in analog outputs (A and B) at terminal block
I/O
, which is
designated as
Slot 0
. Additional analog outputs may be added to the Model GM868 by installing an
Analog Outputs
Option Card in one (or more) of the six expansion slots. Each option card contains four analog outputs, which are
designated as A, B, C and D. Both the zero-point and full-scale values for each output must be calibrated. After
calibrating the outputs, which have a resolution of 5.0
A (0.03% full scale), their linearity should be tested.
Note:
The zero point of the analog output may be set for either 0 mA or 4 mA. However, the calibration procedure
always uses the 4 mA point, as the meter will extrapolate this value to obtain the 0 mA point.
Prepare for the calibration procedure by connecting an ammeter to the desired analog output of
Slot X
, as shown in
Figure 1 below and Figure 2 on the next page. Refer to the menu map in Figure 6 on page 15.
Figure 1: Ammeter Connection for Slot 0 (Output A)
Terminal Block I/O (Slot 0)
B
RTN
B
SIG
A
RTN
A
SIG
4
1
+
-
Ammeter
Summary of Contents for DigitalFlow GM868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 9: ...DigitalFlow GM868 Service Manual vii Contents...
Page 10: ...Contents viii DigitalFlow GM868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Page 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Page 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Page 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Page 71: ......