Chapter 1. Calibration
12
DigitalFlow™ GM868 Service Manual
1.7
Testing the Alarm Relays (cont.)
Note:
The procedure for testing
Alarm Relay A
is identical to that for testing
Alarm Relays B and C
. However, make
sure that the ohmmeter is connected to the desired normally-open or normally-closed contact of the
currently selected relay. See Figure 4 on page 11 for the correct pin numbers on the option card terminal
block.
4.
Pressing
[F1]
(CLOSE)
should yield an ohmmeter reading of about zero. Pressing
[F2]
(OPEN)
should yield an
infinite ohmmeter reading.
5.
Press
[F3]
to
EXIT
.
6.
Repeat the above procedure until both the normally-open and normally-closed contacts for all three alarm
relays have been tested. Then proceed to
Procedure Options
below.
Procedure Options
You have completed testing the alarm relays. Do one of the following:
• To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
• To leave the Calibration Menu, press
[F4] (EXIT)
twice.