Chapter 1. Calibration
8
DigitalFlow™ GM868 Service Manual
1.5.2
aLOW Option = [F1]
1.
Enter the low reference value and press the
[ENT]
key.
2.
Press
[F1]
to store the current low reference value or press
[F2]
to cancel the entry. In either case, the
ANALOG
INPUT
prompt will reappear.
3.
Press
[F2] to proceed to the next section.
1.5.3
aHIGH Option = [F2]
1.
Enter the high reference value and press the
[ENT]
key.
2.
Press
[F1]
to store the current high reference value or press
[F2]
to cancel the entry. In either case, the
ANALOG
INPUT
prompt will reappear.
3.
Proceed to
Procedure Options
below.
1.5.4 4 mA Option = [F3]
1.
Connect the 4 mA current source to the currently selected analog input, as shown in Figure 3 on page 6.
2.
Press
[F1]
to store the current 4 mA value or press
[F2]
to cancel the entry. In either case, the
ANALOG INPUT
prompt reappears.
3.
Press
[F4] to proceed to the next section.
1.5.5 20 mA Option = [F4]
1.
Connect the 20 mA current source to the currently selected analog input, as shown in Figure 3 on page 6.
2.
Press
[F1]
to store the current 20 mA value or press
[F2]
to cancel the entry. In either case, the
ANALOG INPUT
prompt reappears.
3.
Proceed to
Procedure Options
below.
Procedure Options
You have completed calibration of the analog output. Do one of the following:
• To calibrate another input, return to Step 3 in
Accessing the Calibration Menu on the previous page.
• To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
• To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Summary of Contents for DigitalFlow GM868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 9: ...DigitalFlow GM868 Service Manual vii Contents...
Page 10: ...Contents viii DigitalFlow GM868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GM868 Service Manual 16...
Page 38: ...Chapter 3 Diagnostics 28 DigitalFlow GM868 Service Manual no content intended for this page...
Page 66: ...Warranty 56 DigitalFlow GM868 Service Manual...
Page 70: ...60 DigitalFlow GM868 Service Manual no content intended for this page...
Page 71: ......