Document #: 020-17475-00 Revision: E
Page #: 25 of 63
O.
MAINTENANCE & CARE OF SUPERCONDUCTING DRIVE UNIT
The LevMixer
®
Superconducting Drive is designed to operate with minimal
maintenance. However, to minimize wear it is recommend that the drive be
unplugged if it is not in use for more than 24 hours.
The LevMixer
®
drive unit can operate continuously for up to 45 days.
To recharge superconductors the machine must be powered off for at least 25 minutes.
When the unit is started with cold superconductors (short term power off), excessive
noise and vibration are possible within first 15 min of operation. This is normal and is not
considered a malfunction.
Recipes can be transferred between LevMixer units, including the whole recipe library,
using external media (micro SD card).
IMPORTANT!
Update/restore of software resets the runtime counters to zero.
IMPORTANT!
Update/restore of software results in recipes content which is set to factory
installed values (run time and pause time are zero, speed = 20 RPM).
If user defined library content needs to be transferred back in the unit after service is
done it is recommended to save the recipe library on a micro SD card prior to
conducting software service.
O.1 PREVENTIVE MAINTENANCE
Periodical maintenance is recommended to keep the drive unit in reliable working
condition. Wearing of moving parts can be monitored through the elapsed run time
counter in the PLC. Run time information is accessible for viewing on the touch screen
through the setup menu.
Recommended preventive maintenance procedures are listed in Table 2.
Table2. Preventive maintenance procedures
Description
Frequency
Spare parts involved
Who performs
Filter inspection. Replace
if deposits on the filter
media are noticeable.
Every 1000
system run
hours
Replacement Filter Tissue
LT-SVSP327
User
Timing belt replacement Every 9000
motor hours
Timing belt LT-SVSP003
Service
Battery replacement
6 years
Li battery CR2477
Service