4.
Press the main power button on the control box; the button will illuminate when
activated. The cryocooler and touch screen will activate. A PLACE CHARGER
IMMEDIATELY notification will be displayed on the touch screen to remind the user to
complete step 3.
5.
Use the touch screen to acknowledge that the charger is in place to proceed to the
power up screen.
6.
The charging procedure will start automatically. A blinking CHARGE IN PROGRESS will
display on the power up window status bar. Certain initial conditions may prevented
the system from initiating charging after power on. A notice will display on the screen if
an issue exists. See details of power up logic in Section P.3.
7.
Press the blinking NONE button on the Power up screen. When the entry screen opens,
select the type of impeller: use up and down arrows to set four-magnet impeller or six-
magnet impeller type according to matching the impeller type within the mix bag to
be used. Press ENTER. The selected impeller type will now be shown on the power up
screen and a charging timer will indicate the elapsed charging time in the charging
status area (bottom row of power up screen).
8.
Superconductor charging will take approximately 35 minutes. After charging is
complete a blinking READY button will appear on the screen and the CHARGED notice
will be displayed instead of charging time. Charging is complete assuming the charger
was placed on the levitation head prior to powering on the unit.
9.
Proceed to Main Menu window by pressing the Main Menu button.
10.
Remove the magnetic charger and replace the protective shield on the charger. IT IS
CRITICAL THAT THE CHARGER BE PLACED ON THE LEVITATION HEAD FOR THE ENTIRE 35
MINUTE CHARGING PERIOD.
Always put the protective shield back on the magnetic
charger when charging procedure is complete.
MAKE SURE
charger can rotate free being installed in recess of levitated head. It should
sit loose enough to get right orientation.
DO NOT
use alternative spacing/separating devices when charging the drive.
DO NOT
disturb or remove the charger from levitation head until charging is completed.
Use only the magnetic charger provided in the kit.
Power off for more than 25 minutes is required to reset the superconductors.
While drive unit is charging, you can continue with Dolly – Tank Assembly
(Section 7).