Page
40
of
48
Double jacket over-pressure
protection
Interlock using disposable pressure sensor for measurement of the double
jacket internal pressure
Emergency stop
2 emergency stop buttons, 1 on the front and 1 on the left side of the
iCELLis 500+ control system Top Assembly.
Power failure data protection
Integrated UPS for the safe shut down of the computer only. A power
supply UPS is highly recommended
Software platform
Wonderware
♦
ArchestrA system platform
Data storage capacity
2 x SSD of 240 Go (RAID1 for redundant data storage)
Electronic records &
electronic signatures
Compatible with FDA 21 CFR Part 11 and Eudralex Annex 11
Automation design
Developed and tested in accordance with GAMP5 guidelines
Data export/Network
communication
Integrated OPC server for real-time data monitoring. Access to SQL
database for historical data storage
HMI
19 in. touch panel & USB interface
Remote operation
Available via Windows
♦
Remote Desktop Protocol
Cleaning solutions
Recommended cleaning solutions for wiping exposed surfaces of
the hardware:
•
Water
•
70% (v/v) ethanol in water
•
Isopropyl alcohol (IPA)
Recommended preventative
maintenance interval
12 months
Storage temperature
4 °C – 40 °C
Maximum relative humidity in
storage
80% RH, non-condensing
Operating temperature
18 °C – 25 °C
Relative humidity during
operation
10 – 70% RH, non-condensing
Noise pressure level
< 85dBA