WARNING!
Risk of injury to personnel and/or damage to equipment!
The positioning of the accessories pumps (optional FILL and DRAIN pumps) must be done by two
people due to the weight of the pump.
WARNING!
Risk of injury to personnel and/or damage to equipment!
Accessories pumps (optional FILL and DRAIN pumps) must be properly placed and stability of the
pump must be checked before operation.
CAUTION!
Risk of entanglement and crush injury when using peristaltic pumps!
Do not allow fingers or clothing to contact moving parts of the peristaltic pumps.
Keep peristaltic pumps cover closed when pumps are operating.
User must stop the peristaltic pumps before installing and removing tube from peristaltic pump heads.
ALERT!
Risk of damage to equipment!
Only the iCELLis 500+ bioreactor provided by Pall should be used with the iCELLis 500+
control system.
ALERT!
Risk of damage to equipment!
Never block the cooling fan outlets and cooling air inlet on the iCELLis 500+ control system when the
iCELLis 500+ control system is powered.
ALERT!
Risk of damage to equipment!
Only exhaust filter heater provided by Pall should be used with the iCELLis 500+ control system.
ALERT!
Risk of damage to equipment!
The iCELLis 500+ control system should never be moved when there is liquid in the bioreactor.
ALERT!
Risk of damage to equipment!
NEVER restrict flow of gas through the exhaust lines of the iCELLis 500+ bioreactor (gas out line).
ALERT!
Risk of damage to equipment!
The iCELLis 500+ control system is equipped with load cells under the docking station for bioreactor
weight measurement. Load cells must be secured using supplied locking plates during control system
transport and handling in order to avoid damage.
Locking plates must be removed prior to use of the system.
WARNING!
Magnetic field! Risk of injury to personnel!
The iCELLis 500+ control system and iCELLis 500+ bioreactor contain strong magnetic elements.
Persons with certain metallic, electronic, magnetic, mechanical implants, devices, or objects may not
enter this area.
ALERT!
Risk of damage to equipment!
Do not tilt the iCELLis 500+ bioreactor during the draining of the double jacket while it is on the control
system. The load cells can be damaged.