WARNING!
Bended tubing! Risk of injury to personnel and/or damage to equipment!
Avoid sharp bends in the fluid tubing of the TCU, and maintain a sufficient distance from surrounding
walls. Make sure that the tubing is securely attached.
CAUTION!
Risk of injury to personnel and/or damage to equipment!
When the temperature control is active or when the system is cooling down, the TCU media pre-heater
and filter heater may become extremely hot. Therefore, exercise particular caution when handling.
ALERT!
Risk of damage to equipment!
Never block the cooling fan outlets on the TCU.
There are two different TCU configurations available with the iCELLis 500+ control system. One
configuration must be supplied with 230Vac - 50Hz, and the other must be supplied with 120Vac - 60Hz.
The configuration of the TCU must be specified according to the client requirement during the purchasing
of the iCELLis 500+ control system.
The TCU is packed in the crate of the accessories and documentation.
The TCU has been tested to operate with up to 10 meters of tube between the TCU and the bioreactor
(10 meters of tube from TCU to DJB In and 10 meters of tube from DBJ Out and TCU). The TCU must
be placed in a range less than 10 meters around the iCELLis 500+ bioreactor. The TCU is IP21 rated,
please refers to the operating manual of the TCU for cleaning.
Unpack the TCU. Place the TCU on the same floor level as the iCELLis 500+ control system. Connect
the TCU to the iCELLis 500+ control system using the PROFIBUS cable between the Interface panel of
the TCU (Figure 25a and b) and the dedicated connection slot at the rear of the control system (Figure
25c).