Turn the main power switch to ON to power the control system (Figure 27e-f).
Figure 27: iCELLis 500+ control system powering –
a) iCELLis 500+ main power
socket and switch, b) Unscrewing protective cap, c) Removing protective cap,
d) Plugging power cable, e) Locking power cable, f) Switching ON the control system.
5.9 Fitting the Gas Supplies to the iCELLis 500+ Control System
DANGER!
Oxygen causes a fire/explosion risk!
Users are responsible for performing their own safety assessment and obtaining expert advice on the
design of any system used to supply gases to the iCELLis 500+ control system.
RELEVANT ONLY WHEN USING OXYGEN: In the unlikely event of a tubing or connection failure, oxygen
may build up in and around the iCELLis 500+ control system. A means of independently monitoring
oxygen close to the equipment is highly recommended. The monitoring system should be set up to
prevent oxygen levels from reaching 24% or limits recommended by local regulations. In the event of a
rise in oxygen above this level, the supply should be immediately shut down. The iCELLis 500+ control
system can deliver oxygen up to 7 L/min. Oxygen supply to the iCELLis 500+ control system is the
user’s responsibility.