56
REMOVING A PASSWORD CONT'D
7. Press the third (and last) key of the Password. The
Digital Readout will now display “PS-”.
8. Press the
TIP SET
Key.
9. The Calibration Mode has been automatically entered
and the Digital Readout will now display “CAL”. If no
calibration parameters need to be changed, exit
Calibration Mode and enter Temperature Display Mode
(normal operation) by pressing the
TIP OFFSET
Key.
Otherwise, perform calibration now (refer to the
CALIBRATION/SYSTEM SETUP section). When
finished with calibration, press the
TIP OFFSET
Key.
Figure 61. Digital Readout "P S -"
Figure 62. Press Tip Set Key
Figure 63. Digital Readout "CAL"