10.6.3 Parameter 3. DISPL
3
dISPL
3.1
UPdAt
Set the update cycle of display. The update cycle affects the life duration of battery. Even if "off"
is set and the display disappears, the information such as parameter and total value will not be
lost and the flow rate will be surely calculated.
Selectable range: FASt - 1 SEC - 3 SEC - 15 SEC - 30 SEC - OFF (standard setting: 1sec)
If you select "OFF", the display will be automatically turned OFF in 30 seconds after the final
operation, however, if you touch any key, the display will be turned ON again.
NOTE: If the application does not require the high-speed display update cycle, it is
recommended to set the update cycle low.
3.2
GrAPH
The bar chart display of instantaneous flow rate can be ON/OFF.
EnAbLE (enable): ON / dISAbLE (disable): Off
3.3
SPAn
The range span of bar chart of instantaneous flow rate can be set. This display divides the
range span 0 to 100% into ten.
Settable range: 0000.001 - 99,999
NOTE: The decimal point set by decimal point position of instantaneous flow rate of "2.2
dECS" is displayed.
3.4
LIGHt
The lighting time of backlight can be set. If the lighting time becomes long, it affects the life
duration of battery.
Settable range: 0 to 99 sec (if you select 0, the backlight does not light up)
10.6.4 Parameter 4. METER
4
MEtEr
4.1
UnIt
The unit of meter coefficient can be set.
4.2
M.FACt
The meter coefficient can be set.
Settable range: 0.00001 to 99999999
The decimal point set by "4.3 dECS" decimal point position of meter coefficient is displayed.
4.3
dECS
The decimal point position of "4.2 M.FACt" meter coefficient can be selected from the following.
Settable range: 0 - 1 - 2 - 3 - 4 - 5 (0 has no decimal point)