E-126-3-E
11
Fig. 4.7 Dimensions of metal clamp
100 mm (3.94")
103 mm (4.06")
25 mm (0.98")
70 (2.76")
115 mm (4.53")
118 mm (4.65")
7 mm (0.28")
10 (0.39")
72
42
58
22
M5x0.8
68 0.1
63 17
R5
14
14
24
UP 45
R
1
.5
UP 45
R
1
.5
1.50
21.8
56.2
Flat pattern
Bended model
Mounting kit 1
Mounting kit 2
5. EXTERNAL WIRING GLAND
In case of TIIS explosionproof type
(1) This totalizer when used as a flameproof instrument uses the cable gland (pressure-tight packing) method of
IEC harmonized standards for the external wirings. Do not use cable glands other than the provided together
with the instrument. The rubber packing is provided with the product according to the finished external
diameter of cable as shown in the table below. Select and assemble the rubber packing which matches with
the external diameter of your cable.
External dimension of applicable cable
Nominal code of
packing
Internal diameter of
packing
External diameter of
cable
Applicable torque
Pressure tight packing
model code
9
φ9
φ8.1 to φ9.0
18.5N
・
m
KXB-25, 22 ORES
10
φ10
φ9.1 to φ10.0
22.0N
・
m
11
φ11
φ10.1 to φ11.0
25.8N
・
m
KXB-25, 22 ORE
12
φ12
φ11.1 to φ12.0
30.0N
・
m
13
φ13
φ12.1 to φ13.0
33.5N
・
m
NOTE: "Applicable torque" is applied to the gland shown in the assembly diagram below.