5 Troubleshooting
5-24
Service Manual – Osmetech OPTI™ Analyzer
Osmetech OPTI Troubleshooting - 3
Errors during
Calibration
(Meas. & Control)
'Misseat 2'
Message
'Misseat 1'
Message
'Check Pump'
Message
Check Home
Pos. visually
(Pins in 12 & 6
o'clock pos.)
Check top plate to
move freely.
Remove any
debris from SMC
Check DIAG
LEDs LT1 & LT2
toggle
Rerun meas.
using new
cassette
Rerun meas.
using new
Cassette
Was cass.
wiped before
insertion?
Clean Optics &
Upper Plate
(LG1, LG2 free)
Check DIAG
LED L1, L2, L3
with new cass:
1200 - 2800
Run Flowtest
DIAG-Flow
(use new cass.)
Rerun meas.
using new
cassette
Return
Instrument
(RA)
Replace Pump
cart, check for
leaks. Clean
Nipple in SMC
Rerun Flow test
DIAG-Flow
(use new cass.)
Flow Test fails?
NO
YES
Flow test fails?
ok
YES
NO