2 General Description
Service Manual – Osmetech OPTI™ Analyzer
2-5
Data Management
Osmetech
OPTI 1
Osmetech
OPTI CCA
Printout built-in
thermoprinter
3
3
Serial Interface
1x RS 232, 9-pin
SUBMIN D/F
ASCII or
ASTM
ASCII
only
3
Infrared Interface
1XIR (Unidirectional or
Bi-directional (Osmetech
OPTI CCA only))
ASTM
3
Gas Supply
Calibration Gas
14 % O
2
, 6 % CO
2
, balance N
2
,
max. pressure 140 psi (9.7 bar)
Operating Parameters
Sample type
Heparinized whole blood, plasma/serum (pH only), Osmetech
approved quality control materials
Sample input device
syringe, capillary, microsampler
Sample size
Osmetech OPTI 1 80
µ
L
Osmetech OPTI CCA 125
µ
L
Sample input
automatic aspiration
Analysis time
< 2 minutes
Type of measurement
Optical fluorescence and reflectance
Units
conventional, standard international (SI)
Temperature / Humidity
Ambient temperature
15 °C - 32 °C (60 °F – 90 °F) – Osmetech OPTI 1
10 °C - 32 °C (50 °F – 90 °F) – Osmetech OPTI CCA
Measuring chamber
temperature
37 °C
±
0.1 °C (98.6 °F
±
0.18 °F)
(Osmetech OPTI 1 & Osmetech OPTI CCA)
Relative humidity
Altitude
5 - 95 % non-condensing (Osmetech OPTI 1 & Osmetech OPTI
CCA) up to 3048 m (10,000 ft)
Electrical Supply
Mains supply
100 ± 10% VAC to 240 ± 10% VAC
Frequency
50 - 60 Hz
Power consumption
45W max (7W standby)