6 Osmetech OPTI™ Repair Procedures
6-32
Service Manual – Osmetech OPTI™ Analyzer
Figure 27: Printer mechanism
Verification
Go to menu
DIAG2- Printr
and perform a test print.
Remove and re-insert the paper. Make sure the printer auto-feeds the paper upon
insertion.
Analyzer Housing
The main analyzer housing is comprised of two parts, the lower and upper
housing. The assemblies located inside the analyzer can be removed. To gain
access to these assemblies, the analyzer housings must be separated. To open the
analyzer housing, first remove the battery on the left side of the analyzer. Turn
the analyzer over to access the seven (7) screws securing the two housing
sections together. Remove the screws and separate the upper and lower housing
sections by flipping the lower housing open on the side of the carrying handle.
Caution: When turning the analyzer over to access the lower housing section,
place the analyzer on a padded surface to prevent scratches or other damage to
the upper housing section.
Paper feed
Paper guide
Gray lever
Printer shroud