3 Test Programs and Diagnostics
Service Manual – Osmetech OPTI™ Analyzer
3-7
LED Test
The LED test displays the operating voltage of the internal LED’s and allows for
verification of operation and performance. The LED test will display the
following LED’s:
LT1
Cassette seat detect front (SMC Top Cover)
LT2
Cassette seat detect rear (SMC Top Cover)
L1
Front fluid detect LED
L2
Rear fluid detect LED
CD
Cassette detect LED
Memory Test
The
Memory Test
can be selected and the analyzer performs a check of the
system memory. The analyzer displays
MEMORY OK
when the system
successfully verifies the memory.
Software Version
Selecting the
Software Version
menu displays the currently installed software
revision level. Pressing the right arrow key displays the software version for the
optics module.
Reports
With the
Reports
menu, the user can select one of four reports to be printed.
The
Error Report
saves the last 50 system errors and can be printed. When the
number of system errors exceeds 50 the oldest error is deleted and replaced by
the most recent error.
A
Calibration Report
can also be printed which provides a calibration data
summary of the stored patient and QC measurements.
The third report is an
SRC Report
, which provides a summary of all stored SRC
results.
The
Configuration Report
prints out all user selectable settings. A printout of
this report is particularly helpful prior to erasing memory (e.g. software update).