Easy-MCS
™
Multichannel Scaler and MCS-32 (A73-B32) Software
931054C / 1213
110
SET_LLSCA 1.2
SET_PASS_LENGTH 1000
SET_PASS_PRES 0
SET_ULSCA 6.5
RESET
Resets the MCS to the state just after power is applied.
SET_DISCRIMINATOR voltage
Sets the input discriminator threshold to the specified voltage. Voltage has a maximum value
of +3.0 and a minimum value of
!
1.6. Legal values b3.0 and
!
1.6 occur nominally
every 0.0015256 volts. When an invalid value is specified, the value is rounded off to the
closest legal value and a warning message is returned. See also SHOW_DISCRIMINATOR.
SET_DISCRIMINATOR_LEADING
Makes the input discriminator leading (rising) edge sensitive. After this command is issued,
the counters will increment when the input crosses the threshold with a positive slope. See
also SHOW_DISC_EDGE and SET_DISCRIMINATOR_TRAILING.
SET_DISCRIMINATOR_TRAILING
Makes the input discriminator trailing (falling) edge sensitive. After this command is issued,
the counters will increment when the input crosses the threshold with a negative slope. See
also SHOW_DISC_EDGE and SET_DISCRIMINATOR_LEADING.
SET_DWELL dwell
Sets the dwell time to the specified value. Dwell is in units of seconds and exponential nota-
tion is permitted. Dwell has a maximum value of 65535 and a minimum value of 5×10
!9
.
Legal values between the minimum and maximum values may be determined by using the
VERIFY_DWELL command. The value can then be set with SET_DWELL. If an invalid
value is specified, the value is rounded to the closest legal value and a warning message is
returned. See also SHOW_DWELL, ENAB_DWELL_EXT and DISA_DWELL_EXT.
SET_DWELL_ETHLD voltage
Sets the external dwell threshold. The value of voltage can be a floating-point number
between –1.6 V and +3.0 V. Legal values occur between these limits nominally every
0.0015256 volts. When an invalid value is specified, the value is rounded off to the closest
legal value. See also SHOW_DWELL_ETHLD.
SET_INPED number
Sets the input impedance. Number has only 2 valid values, zero and 1. Zero sets the input
impedance to 50
Ω
; 1 sets the input impedance to 1 k
Ω
. See also SHOW_INPED.
Summary of Contents for Easy-MCS
Page 10: ...x...
Page 20: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 10 Intentionally blank...
Page 24: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 14 Intentionally blank...
Page 77: ...931054C 1213 7 KEYBOARD FUNCTIONS 67 Fig 59 MCS 32 Keyboard and Keypad Accelerators...
Page 98: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 88 Intentionally blank...
Page 100: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 90 Intentionally blank...
Page 108: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 98 Intentionally blank...
Page 110: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 100 Intentionally blank...
Page 114: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 104 Intentionally blank...
Page 130: ...Easy MCS Multichannel Scaler and MCS 32 A73 B32 Software 931054C 1213 120 Intentionally blank...