932502G / 0618
7. FIRMWARE COMMANDS AND RESPONSES
Non-USB Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was
in progress. No action was taken.
COMPUTER
No function in the DSPEC 50; included for backward compatibility.
DISABLE_BLRE_AUTOMATIC
Disables the automatic determination of baseline restorer time constant. See also
ENABLE_BLRE_AUTOMATIC, SET_BLRE, and SHOW_BLRE.
DISABLE_ETP
Returns the DSPEC 50 to normal operation if in Enhanced Throughput (ETP) mode. See also
ENABLE_ETP, SHOW_ETP, SET_TIME_ETP, LIST_TIME_ETP, and
SHOW_TIME_ETP.
DISABLE_FAST_COIN
Disables the Fast Coincidence function used with Coincidence or AntiCoincidence gating.
DISABLE_GAIN_STABILIZATION
Stops stabilization of the gain peak while data is being acquired. The gain stabilization
adjustment is held at its current value until either gain stabilization is reenabled with the
ENABLE_GAIN_STABILIZATION command or reinitialized with the
INITIALIZE_GAIN_ STABILIZATION, SET_GAIN_PEAK or SET_GAIN_WIDTH
command. See also SHOW_GAIN_STABILIZATION.
DISABLE_HV
Turns off the high-voltage enable signal of the DSPEC 50. See the section on the bias supply
for more information about the high-voltage output. See also ENABLE_HV and
SHOW_HV.
DISABLE_HV_DIM
Disables the use of the DIM connector on the back of the DSPEC 50 and causes the
DSPEC 50 to suppress SOH information (see SHOW_HV_HEALTH command). Also, the
command disables polling of a standard DIM module (e.g. DIM-POSNAI, DIM-POSGE,
DIM-NEGGE, or DIM-296) by the DSPEC 50.
DISABLE_LFR
Returns the DSPEC 50 to normal operation if the LFR filter is enabled. See also
ENABLE_LFR and SHOW_LFR.
83