DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
THR_POS 00000126811
Positive threshold is 126811.
SHOW_TIME
Reports the time from the battery backed-up system clock in the form:
$Nhhhmmmsssccc<CR>
where hhh represents a 3-digit integer hour (0–23), mmm represents a 3-digit integer minute
(0–59), and sss represents a 3-digit integer second (0–59). See also SET_TIME,
SET_DATE, and SHOW_DATE.
Example Responses:
$N010054017052<CR>
Time returned 10:54:17 (10 h, 54 min, 17 s).
$N020013037050<CR>
Time returned 20:13:37 (20 h, 13 min, 37 s).
SHOW_TIME_ETP
Returns the current Enhanced Throughput Mode protection time.
Response:
TIME_ETP 0000000000012.8
Protection time is set to 12.8 ìs.
SHOW_TIME_START
Reports the time of the last START command in the form
$Nhhhmmmsssccc<CR>
where hhh represents a 3-digit integer hour (0 through 23), mmm represents a 3-digit integer
minute (0 through 59) and sss represents a 3-digit integer second (0 through 59). See also
SET_TIME_START, SET_DATE_START, and SHOW_DATE_START.
Example Responses:
$N010054017052<CR>
Time returned 10:54:17 (10 h, 54 min, 17 s).
$N020013037050<CR>
Time returned 20:13:37 (20 h, 13 min, 37 s).
SHOW_TRIGGER
Tells whether the trigger is enabled or disabled. See also ENABLE_TRIGGER.
Responses:
$IF
Trigger disabled.
$IT
Trigger enabled.
SHOW_TRIGGER_POLARITY
This returns the polarity setting of the input trigger for the sample changer (1 means the
trigger is positive-edged and triggers on a low-to-high transition; 0 means it is negative-edge
triggered and triggers on a high-to-low transition). See also TRIGGER, ENABLE_
TRIGGER, DISABLE_TRIGGER, SHOW_TRIGGER, and SET_TRIGGER_POLARITY.
Responses:
$FNEG<CR>
The HV is set to negative.
$FPOS<CR>
The HV is set to positive.
116