DSPEC 50
®
and DSPEC 502
®
Digital Gamma-Ray Spectrometer User’s Manual
932502G / 0618
ENABLE_OPTI_NEED
Enables the reporting of an error when the DSPEC 50 needs optimizing.
ENABLE_PUR
Enables the Pile-Up Rejection Circuitry.
ENABLE_PZ_AUTOMATIC
Enables the automatic pole zero mode. See also DISABLE_PZ_AUTO and
SHOW_PZ_AUTO.
ENABLE_PZ_NEED
Enables the reporting of an error when the DSPEC 50 needs pole zeroing.
ENABLE_REMOTE
No function in the DSPEC 50; included for backward compatibility.
ENABLE_RENH
Sets the DSPEC 50 to Resolution Enhancer mode. See also SHOW_RENH.
ENABLE_THRESHOLD_AUTOMATIC
Enables automatic determination of the positive and negative thresholds. See also
DISABLE_THRESHOLD_AUTO, SHOW_THRESHOLD_AUTO,
SET_THRESHOLD_NEGATIVE, and SET_THRESHOLD_POSITIVE.
ENABLE_THRESHOLD_FMAN
Enables manual adjustment of the fast channel threshold.
ENABLE_THRESHOLD_SAMPLE
ENABLE_TRIGGER_SPECIAL
No function in the DSPEC 50; included for backward compatibility.
ENABLE_TRIGGER
Enables the Sample Ready connector. Acquisition will then start when an external trigger
occurs on the Sample Ready connector or when the TRIGGER command is issued. See also
TRIGGER, DISABLE_TRIGGER, SHOW_TRIGGER, SET_TRIGGER_POLARITY, and
SHOW_TRIGGER_POLARITY.
ENABLE_ZDT
Sets the DSPEC 50 to ZDT mode. See SET_MODE_ZDT for information on selecting the
ZDT mode.
86