3.3. Step 3: Install the Spectroscopy Application Software
Install MAESTRO or another ORTEC spectroscopy application according to the instructions in
its user manual.
3.4. Step 4: Connect the DSPEC 50 to Your Network
or Computer
NOTE
You can attach DSPEC 50s directly to your computer
either
by Ethernet
or
by USB
connection, but do not mix the two connection types on the same computer or opera-
tional errors could occur. If you switch a DSPEC 50 from one connection type to the
other, you must re-run the MCB Configuration program to reestablish communication
between the DSPEC 50 and the computer.
3.4.1. Ethernet Connection
!
Case 1: LAN With DHCP Server — Dynamic IP Addressing
Most users will connect one or more DSPEC 50s, via the rear-panel
Ethernet
port, to a stan-
dard LAN. This is the simplest connection method: Just attach the DSPEC 50 to your net-
work with a standard Cat 5 (RJ45) Ethernet cable, power the DSPEC 50 on, wait 10–15
seconds for the network’s DHCP server to assign a dynamic IP address to the DSPEC 50,
and you’re ready to run the MCB Configuration program (Section 3.5).
In addition to these
networked DSPEC 50s, you can attach DSPEC 50s directly to your computer
either
by
Ethernet
or
by USB connection
, as described in the next two paragraphs and Section 3.4.2.
!
Case 2: LAN Without DHCP Server or Standalone Laptop/Computer With LAN
Adapter — Dynamic IP Addressing
If you (a) connect multiple DSPEC 50s to a network that does not have a DHCP server, or
(b) have an isolated computer with a LAN adapter, and wish to attach multiple DSPEC 50s to
the computer via a powered Ethernet switch, Windows will auto-assign IP addresses to each
DSPEC 50. To do this, attach the DSPEC 50s to your network (or to the standalone com-
puter’s Ethernet switch) with a standard Ethernet cable, power the DSPEC 50s on, wait
3–5 minutes for the IP addresses to be assigned, and you’re ready to run the MCB Config-
uration program (Section 3.5).
!
Case 3: Static IP Addressing
(a) If for any reason the network DHCP server or the host computer does not auto-assign an
34