42
Output: 9-Pin Female D-sub Connector
Failed part output. Closes when measurement is out of software specified range.
SPDT Relay, Resistive load Max: 200mA
Max Ratings: 30VDC @ 1A, 125VAC @ 0.1A
Internal parts: Omron G6E, connected internally with 28AWG ribbon cable
Pin
Signal
1
*
Anal (0-10V / 4-20mA)
2
Not Used
3
Not Used
4
GND (system ground)
5
Ra_Limit (Closed if Out of Spec)
6
Analog_Out Ground (0-10V / 4-20mA)
7
Not Used
8
Not Used
9
Ra_Limit (Closed if Out of Spec)
* NOTE:
Jumpers JP2 and JP3 on top board inside of control box must be set to configure 0-10V voltage
or 4-20 mA current output. Set JP2 and JP3 to “C” for 4-20 mA current output or to “V” for 0-10V voltage
output.
1)
Output Schematic Ra_Limit
a)
Normal State – Surface Within Ra_Limit – in Spec
b)
Surface Outside Ra_Limit – out of Spec
2)
Output Schematic Analog_Out
a)
0-10V
b)
4-20mA
Pin 9
Pin 5
Pin 9
Pin 5
Pin 1
Pin 6
0 Volt @ LOW_ANALOG
10 Volts @ HIGH_ANALOG