46
Reset Command
This selection should only be used if the user experiences difficulties with files and software in the
controller. Scroll to FILES: >Delete on the control box with the scroll (>) push button and push the Select
push button. The LCD displays the following screen.
Reset to factory
files? >y/N
Scrolling to highlight “Y” and pushing the Select push button will present the following display:
ALL data will be
deleted! >y/N
Scrolling to highlight “Y” and pushing the Select push button will delete data, configuration, and setup files
that have been added to the control box and return it to factory setups and configurations.
Exit Command
This selection exits the file management section of the system and returns to the Main Screen.
Setup Files
Every Lasercheck comes with setup files, GRIND_ST.SET, UNCAL_R.SET, and UNCAL_S.SET.
GRIND_ST is used to measure surfaces that have been ground. UNCAL_R and UNCAL_S are setup files
that are used to collect data from rough and smooth surfaces when creating a new setup file.
Creating Setup Files
Included with the software is an Excel spreadsheet program titled “Setup File and Calibration
Template.xls”. This program is found in the C:\Program Files\Lasercheck\Calibration directory created by
the installation program. It is designed to calibrate the Lasercheck for specific processes. The spreadsheet
program also integrates the calibration in Lasercheck setup files.
Detailed instructions on how to perform calibrations and use this Excel program are in the appendix of this
manual under the section titled “Appendix - Calibration Procedures”. The three setup files that the
spreadsheet creates can be renamed as long as the name is not longer than eight characters and its three
character extension, “.SET”, remains. These setup files need to be transferred to the control box; see the
Receive Command in the section on File Management.