8-3
8. Specifications
Output Section
Analog output format
Voltage output 0
±
10V/F.S
Scale
Attenuation setting of the full scale is possible from 0.1V to 10V in 0.1V
steps.
Response
16ms to 64s depending on the time constant setting
OFF is provided only when the TS-0321 option is mounted.
Accuracy
When combined with the detector (with 1s average value)
N-0 compensation used:
±
0.2% / full-scale
N-0 compensation not used:
±
0.5% / full-scale
Temperature drift (TS)
±
0.01% / full scale /
°
C
Appropriate load resistance
10k
Ω
or higher
Summary of Contents for TS-3200A
Page 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Page 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Page 96: ...4 30...
Page 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Page 150: ...6 8...
Page 151: ...7 1 Chapter 7 Troubleshooting...
Page 154: ...7 4...
Page 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Page 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Page 170: ...8 16...