3.5.1 Manual Setting of N-0 Compensation Values
Example: Setting N-0 compensation value for the CW direction
Prepare the following table, rotate the detection shaft using a motor, etc., then record the torque display
value for each rpm.
Note: Like zero compensation, enter display values, with the decimal point removed, in the
count column. This also applies to rpm display values.
Then, plot values using the horizontal axis for the rpm and the vertical axis for the torque display value to
determine compensation points.
The detector can be thought to have the frequency characteristic shown by the broken line shown below.
When approximation is made linearly using 4 compensation points, Nos.1 to 4 are selected. In the above
example, when more data are taken around 2000r/min where the curve largely changes, compensation
values can be set more finely resulting in more accurate measurement.
1 %
-0.5 %
+10
-5
+5
-10
3,000
4,000
2,000
1,000
5,000
6,000
No.3
No.2
No.1
+8
+6
No.4
Point
Rpm
Compensation
value
No. 1
500
-8
No. 2
1800
+6
No. 3
4000
-5
No. 4
6000
-5
Rpm (r/min)
Count
0
500
-10
1000
0
2000
5
3000
0
4000
-5
5000
-5
6000
-5
3. Basic Operations
Summary of Contents for TS-3200A
Page 1: ...TS 3200A DIGITAL TORQUE METER INSTRUCTION MANUAL ONO SOKKI CO LTD...
Page 21: ...Chapter 2 Before Use 2 1 Name and Function of Each Section 2 2 Detector Connections...
Page 96: ...4 30...
Page 143: ...6 1 Chapter 6 Menus 6 1 MENU Key 6 2 CAL Key 6 3 TRQ ZERO Key...
Page 150: ...6 8...
Page 151: ...7 1 Chapter 7 Troubleshooting...
Page 154: ...7 4...
Page 167: ...8 13 Panel cutout dimensions 4 M5 370 405 100 1 0 50 0 5 4 2 6 34 301 12 8 Specifications...
Page 169: ...8 15 4 2 6 34 301 12 8 Specifications...
Page 170: ...8 16...