4-4
Sect
ion 4
APPLICATION SETTING EX
AMPLE
S
ZS-MDC
User’s Manual
Section 4
Measuring the Thickness of Multiple Points (sandwiched thickness)
Press MENU key - ESC key.
9.
To set the expression for calculating the dif-
ference in the thickness of the 3 locations
for TASK4, select [TASK4].
[SENSING]-[CALC]-[CALC]-[MAX-MIN]
10.
Set task 1 to task3 for calculating the differ-
ence to ON, and other tasks to OFF.
• To output the judgment result of the operation
[I/OSET]-[I/O SET]-[OUT]
11.
Select [TASK4] so that the judgment on the
thickness difference can be output.
• To linear-output the operation result
[I/O SET]-[I/O SET]-[ANALOG]
11.
Select [TASK4] so that the thickness differ-
ence can be linear-output.
The calculation results of each CH can be batch-acquired if you use the communication command.
▲
1
TASK1
3
TASK3
2
TASK2
4
TASK4
▲
1
TASK1
3
TASK3
2
TASK2
▲
1
TASK1
3
TASK3
2
TASK2
4
TASK4
▲
1
TASK1
3
TASK3
2
TASK2
4
TASK4