With trolley (optional):
Properly connect the
CPC 100
and
CP TD1
grounding terminals to the
trolley’s ground bar. Connect the ground bar to earth.
2. Switch off the
CPC 100
at the mains power switch.
3. Connect the
CP TD1
’s "BOOSTER IN" to the
CPC 100
’s "EXT. BOOSTER"
with the booster cable.
4. Connect the
CP TD1
’s "SERIAL" to
CPC 100
’s "SERIAL" with the data
cable. This cable also provides the power supply for the
CP TD1
.
5. Make sure that all cable connectors are clean and dry before being tightly
connected.
6. Pull out the measuring cables from the cable drum and connect the test
object to the
CP TD1
’s measuring inputs IN A and IN B.
7. Connect the high-voltage cable from the test object to the
CP TD1
’s high-
voltage output.
•
At the
CP TD1
, press the high-voltage cable’s plug to the connector tightly
and turn the screw cap manually without using any tools until you feel a
mechanical stop. If you notice a rough-running of the screw-cap, clean the
screw thread and use a lubricant (vaseline recommended).
•
Insert the yellow banana plug (the high-voltage cable’s grounding) into the
respective plug socket.
•
At the test object, insert the high-voltage cable’s plug carefully until you feel
a "click" position. Now they are locked. Confirm this by trying to pull them
out. This should not be possible now.
•
The high-voltage cable is double-shielded and therefore safe. However, the
last 50 cm (20 inch) of this cable have no shield. Therefore, during a test
consider this cable a life wire and due to the high-voltage life-hazardous!
8. Connect the
CPC 100
to the mains power supply using the provided cable.
Connect the
CPC 100
grounding terminal to
substation ground.
Summary of Contents for CP TD1
Page 1: ...CP TD1 Reference Manual...
Page 14: ...CP TD1 Reference Manual V 1 44 14...
Page 20: ...CP TD1 Reference Manual V 1 44 20...
Page 32: ...CP TD1 Reference Manual V 1 44 32...
Page 38: ...CP TD1 Reference Manual V 1 44 38...
Page 84: ...CP TD1 Reference Manual V 1 44 84 Figure 6 14 Frequency scan for H L f 5 kV...
Page 114: ...CP TD1 Reference Manual V 1 44 114...
Page 128: ...CP TD1 Reference Manual V 1 44 128...
Page 154: ...CP TD1 Reference Manual V 1 44 154 Figure 12 12 Diverter switch for 110 kV 40 MVA...
Page 184: ...CP TD1 Reference Manual V 1 44 184...
Page 186: ...CP TD1 Reference Manual V 1 44 186...