59
Capacitance and Dissipation Factor Measurement
For the separate measurement of all capacitors a so-called guard technique is
necessary. The single capacitors connected to guard are energized but not
measured.
In Figure 5-22 a block diagram of the
CP TD1
is shown with the guard
connection and measuring input. In the example case C
1
, C
2
and C
3
are
capacitors, connected to the
CP TD1
. C
1
is connected to input A, C
2
is
connected to input B and C
3
is connected to ground. All three capacitors are
energized. Only C
1
is measured, because the relay matrix only connects C
1
to
the measuring input (instrument), whereas the currents through C
2
and C
3
are
bypassed. C
2
and C
3
are connected to the foot-point of the HV transformer
(GUARD).
Figure 5-22
CP TD1
block diagram with GUARD and measuring input
To get more familiar with this technique we want to measure C HL, C HT and C
H of Figure 5-21. The high-voltage winding is connected to the test voltage
(high-voltage output of the
CP TD1
), the low-voltage winding is connected to IN
A and the tertiary winding is connected to IN B.
T
Tertiary winding
C T
Cap. T to Ground
C H-T Cap. H to
T
IN A
IN B
C
1
C
2
C
3
UST-A
C = C
1
Guard
Measuring input
Summary of Contents for CP TD1
Page 1: ...CP TD1 Reference Manual...
Page 14: ...CP TD1 Reference Manual V 1 44 14...
Page 20: ...CP TD1 Reference Manual V 1 44 20...
Page 32: ...CP TD1 Reference Manual V 1 44 32...
Page 38: ...CP TD1 Reference Manual V 1 44 38...
Page 84: ...CP TD1 Reference Manual V 1 44 84 Figure 6 14 Frequency scan for H L f 5 kV...
Page 114: ...CP TD1 Reference Manual V 1 44 114...
Page 128: ...CP TD1 Reference Manual V 1 44 128...
Page 154: ...CP TD1 Reference Manual V 1 44 154 Figure 12 12 Diverter switch for 110 kV 40 MVA...
Page 184: ...CP TD1 Reference Manual V 1 44 184...
Page 186: ...CP TD1 Reference Manual V 1 44 186...