Commissioning and operation
Series TCL-3M165E2
- 24 -
4.3.4.3 Filling the micro bath *
IMPORTANT! Pay attention to § 4.3.4.2 "Information on filling amounts"!
When filling, leave enough room for thermal expansion, displacement by sensors and
level rise due to stirring.
Unscrew the cover of the tank/ tub insert.
Insert the tub insert in the tank using the sleeve exchange tool
(only for tub insert).
Place the magnetic stirrer in the tank / tub insert.
Insert the sensor cage.
Insert the test specimens in the sensor cage.
This accounts for the volume of the sensors to be
tested.
Fill the calibration liquid in the tank / tub insert.
Leave sufficient reserve space for an additional rise in
the level.
If necessary, remove the test specimens again.
Screw the work cover onto the tank and insert the sensors through the work cover into
the tank or tub insert.
*
1
For tubs already filled, some step are not required.