Testing process / Calibration
Series TCL-3M165E2
- 46 -
6 Testing process / Calibration
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
IMPORTANT! Allow the calibrator to warm up before first calibration!
Allow the calibrator to warm up for at least an hour before the first calibration. This can
be achieved during the calibrator approaches the first testing point, for example.
Before the testing process:
Before the start of the testing process, check whether
the instructions regarding the installation site and the operational position have been
complied with (
§ 4.1).
the electrical connections have been made correctly (
§ 4.2).
the inside of the block and the surface of the adapter sleeve or the calibration insert are
dry. You can safely remove existing ice or condensation water by heating to above
100 °C (212 °F).
the correct measuring insert has been selected for the test task (
§ 3.3).
the test specimen is securely fixed in the calibrator.
the calibrator has sufficient structural stability.
Starting the testing process:
IMPORTANT! Parameter fields disabled!
During the testing process, all the parameter fields, with the exception of the Start / Stop
slider and the alarm signal, are disabled.
•
Push the button of the start / stop slider completely to the right.
The testing process is started. Information on the
testing process and the temperature progression
is displayed in the window.
Information on the testing process is displayed in the
display range (
p. 47).
Testing process finished:
A dialog box with the message that the test task was successfully
completed is displayed.
•
Tap "OK" to close the dialog box.
•
Push the button of the start / stop slider completely to the left.
The main window is displayed.