Series TCL-3M165E2
Construction, Function and Test Task
Technical changes reserved
- 13 -
3.2 Function
IMPORTANT NOTICE!
For your safety, manual operation of the calibrator, i.e. directly starting it up and its
remaining at a certain temperature, is not possible.
The testing procedure is always started with a testing task (
§ 3.3).
This ensures that the calibrator is always started with a defined "Behavior at test end"
(
p. 25).
The calibrators can be used with dry block, infrared, surface or micro bath function.
The function is determined by the calibrator type and the used measuring insert.
Function
Type
Dry block
Infrared
Surface
Micro bath
Measuring insert: Adapter sleeve
Infrared insert
Surface insert
Calibration liquid
TCL-3M165E2
Method of working:
Adapter sleeves or calibration inserts that are suitable for the temperature sensors or
temperature measuring instruments to be tested are inserted in the calibrator
(
§ 4.3 "Preparing of the calibrator").
They give the test specimen a secure fit and ensure optimum heat transfer.
Once all the preparations have been carried out, the calibrator can be switched on (
§ 4.4).
The calibrator heats or cools the metal block or tank to the set temperature. As soon as this
temperature has been reached and is stable, the calibrator signals accordingly.
The calibration of the test specimen can now be carried out. Then, the next test point is
approached or the testing procedure is ended.
Functionality:
The measuring inserts (
p. 17) of the calibrator have different characteristic values, which
were determined in our laboratory through calibration. These characteristic values are
defined at the factory and stored in the calibrator as a protected function.
To be able to use a measuring insert, the relevant "Functionality of the calibrator" (
§ 5.4)
must be selected.
IMPORTANT! Protected list entries!
The functions, test specimens and test tasks marked with the symbol are factory-
defined. They can be neither deleted nor configured.
They serve as templates for your own functions, test specimens and test tasks, which
are marked with the symbol.