30 |
P a g e
8.6
Trigger Command
The Trigger command allows initiating an action on the smart sensor making use of the options provided by the
Trigger register at register location 0x26. If no option is provided a Trigger Log sequence will be performed, forcing
a reading to be taken and saved to the event memory.
T(rigger) <{Add}> <R(eset) | F(actory reset) | P(ower reset) | W(passWord reset) | C(lear log) | S(ample) | L(og)
Action
Trigger Register, value
Usage
R(eset)
Trigger Register =
Trigger Value 0x0004
Reset device forcing re-enumeration of
sensor mix
F(actory reset)
Trigger Register =
Trigger Value 0x0005
Forces a factory reset which clears all user
set up and logged information
P(ower reset)
Trigger Register =
Trigger Value 0x0006
Performs a user reset that is treated as a
power on reset which includes logging the
event in the event log.
W(password reset)
Trigger Register =
Trigger Value 0x0007
Clears password and data log memory but
does not clear configuration information.
C(lear log)
Trigger Register =
Trigger Value 0x0003
Clears the event log
S(ample)
Trigger Register =
Trigger Value 0x0100
Forces a sampling of the sensor data. The
data
is not
written to the event log. The
display will show the current values.
L(og)
Trigger Register (0x26),
Trigger Value 0x0300
Forces a sampling of the sensor data and
the information
is
saved to the event log.
The display will show the current values.