18 |
P a g e
7.3.10.2
Sub GHz Radio Options
The Radio Option byte provides additional configuration control and is radio specific.
Sub GHz Radio Configuration
15
14
13
12
11
10
9
8
Reserved
Band
Omega Sub Ghz (915 Mhz)
0
0
1
Omega Sub Ghz (816 Mhz)
0
1
0
Reserved
Sub Ghz High Power Enable
X
7.3.10.3
Sub Ghz Power
The Sub GHz Power level bit enables the Power Amplifier on the Sub Ghz radio assembly.
7.3.11
System Status
The System Status register provides information on the IF-006 connection to an attached probe. Some
information is duplicated in the Extension Space status register. It is recommended that the IF-006
Configuration Status register is used for system configuration tools.
System Status Register
7
6
5
4
3
2
1
0
Password
Valid
(PV)
Enumerate
Probe
(EP)
Power
Fault
(PF)
Passwords
Active
(PA)
User PW
Pending
(UPP)
Password
Error
(PE)
Auth.
Error
(AE)
Probe
Attached
(PR)
15
14
13
12
11
10
9
8
Device
Ready
Bootstrap
Pending
7.3.11.1
Probe Attached
The Probe attached (PR) bit is set when a smart probe is detected by the IF-006. The IF-006 will
periodically check the ability to access the probe and will clear this bit if the probe is detached. This
status information is also available in the IF-006 configuration
status
register.
7.3.11.2
Authentication Error
If the Authentication Required control bit is set both the attached probe
and
the IF-006 execute a secure
authentication sequence using a shared Private Key and 128-bit AES encrypted handshaking. If the
authentication process fails, the Authentication Error (PE) bit is set and data cannot be accessed on the
attached probe.
7.3.11.3
Password Error
If the internal IPW value does not match the non-blank UPW of the probe the Password Error (PE) bit is
set and data will not be available to the radio interface..
7.3.11.4
User Password Pending
The User Password Pending bit is set during enumeration if the attached probe UPW is non-blank. It is
cleared by writing the value UPW string thru the USB interface.
7.3.11.5
Power Fault
A power fault was detected on the external probe. The bit will be cleared when the overload condition
has been removed.
7.3.11.6
Enumerating Probe
The Enumerating Probe (EP) bit is set while the interface is in the process of enumerating the attached
probe.