8.9
Serial Command
The Serial command allows setting specific operating characteristics of the Serial (USB) interface. If no option is
provided the current settings for the selected characteristic are provided.
S
(erial) <
B
(audrate) = 9600 | 19200 | 38400 | 115200> <
P
(arity) = E(ven) | O(dd) | M(ark) | N(one)>
<
S
(top) = 0 | 1> <
D
(ata) = 7 | 8> <
R
(eset)>
S
Baudrate = 115200, Parity = Even, Data = 8, Stop = 1
Multiple options may be set on the same command line in any order. The updated configuration will be shown
using the current serial configuration and then all changes are applied at once.
Serial BR=38400, Stop = 1, Data=7 Parity = Odd
Baudrate = 38400, Parity = Odd, Data = 7, Stop = 1
<changes are applied, terminal configuration must be changed to match new configuration>
Characteristic
Options
Usage
Baudrate
9600, 19200, 38400,
115200
Serial Baudrate = 38400
Parity
Even
, Odd, Mark, None
S P=None
Stop
1
, 2
S S=2
Data
7,
8
Serial Databits = 8
Reset
--
Resets serial configuration to 115200,
Even, 8, 1