4

Several typical applications of the DPF500 are shown below. Please observe that the various pulse inputs and power 

options may be intermixed in many ways to solve common applications. The isolated pulse output may be freely 

used so long as proper polarity is observed.

Caution

: When 4-20 mA loop option is provided, the power wiring to the loop power option should always be to 

terminals (+) 12 and (-) 11. Accidental wiring to (+) 12 and (-)6 should be avoided since excessive current 

fl ow may result.

Caution

: The magnetic pickup input and contact closure input require isolated sensors for proper operation. Ac-

cidental connections to earth may result in erroneous operation of the analog output and/or excessive 

current fl ow.

Caution

: Accidental connections from circuit common (3 or 6) to earth or terminal (11) may result in erroneous op-

eration of the analog output and/or excessive current fl ow.

2.4 Wiring

2.5 Typical Applications / Wiring Diagrams

+

+

-

1

 Mag Input 1

Mag Input 2

Shield/GND

 Reset Input

Contact Input 

Common/ DC In (-)

1

-

To Shield

Pin 3

Flowmeter

with DC Pulse Output

+

-

Strip Chart

Recorder

24 VDC

Power 

Supply

+

-

ISOLATED INPUT / 4-20mA LOOP POWERED

NOTE: 

Separate power supplies 

are recommended to isolate output 

from input.

If isolation is not a concern, use a 

single power supply

4-20mA  (+)/DC In (+) 

12

4-20mA  (-) 

11

Opto Input (+) 

10

Opto Input (-) 

9

Opto Out (+) 

8

Opto Out (-) 

7

7

8

9

10

11

12

24 VDC

Power 

Supply

PSU-24B

PSU-24B

Remote 

Reset Switch

1 Mag Input 1

Mag Input 2

Shield/GND

 Reset Input

Contact Input 

Common/ DC In (-)

Turbine Meter  with

Mag Pickup

1

MAG INPUT / DPF5-BP POWERED

4-20mA  (+)/DC In (+) 12

4-20mA  (-) 11

Opto Input (+) 10

Opto Input (-) 9

Opto Out (+) 8

Opto Out (-) 7

7

8

9

10

11

12

DPF5-BP PLUG-IN

CONNECTOR

DPF5-BP

Remote 

Reset Switch

15 VDC

Power 

Supply

+

-

1

 Mag Input 1

Mag Input 2

Shield/GND

 Reset Input

Contact Input 

Common/ DC In (-)

Turbine Meter  with

Mag Pickup

1

MAG INPUT / DC POWERED

4-20mA  (+)/DC In (+) 

12

4-20mA  (-) 

11

Opto Input (+) 

10

Opto Input (-) 

9

Opto Out (+) 

8

Opto Out (-) 

7

7

8

9

10

11

12

FPW-15

Flowmeter  with

Switch Closure Output

4-20mA  (+)/DC In (+) 

12

4-20mA  (-) 

11

Opto Input (+) 

10

Opto Input (-) 

9

Opto Out (+) 

8

Opto Out (-) 

7

1

 Mag Input 1

Mag Input 2

Shield/GND

 Reset Input

Contact Input 

Common/ DC In (-)

1

CONTACT INPUT / PULSE OUTPUT / BATTERY POWERED

DPF76 Totalizer

123456

4     1

7

8

9

10

11

12

Summary of Contents for DPF500 Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info DPF500 Series Ratemeter Totalizer Shop online at User sGuide...

Page 2: ...78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056...

Page 3: ...nsions 3 2 4 Wiring 4 2 5 Typical Applications Wiring Diagrams 4 Chapter 3 20 Point Linearization Operation 3 1 20 Point Linearization LIN option 5 Chapter 4 Programming 4 1 Explanation of Display Pro...

Page 4: ...ii THIS PAGE INTENTIONALLY BLANK...

Page 5: ...Optically Isolated Displays Rate Total Simultaneously 5 Digit Rate Display 8 Digit Totalizer Display 4 20mA Analog Output MA option Powered From Internal Battery or 4 20 mA Output Loop MA option 20 P...

Page 6: ...its end of life A descriptor BAT illuminates when the battery voltage falls below this predetermined value The low bat tery detector operates correctly with all power options The battery or batteries...

Page 7: ...emove cover and 4 panel screws Terminals are on bottom side of PC board Panel Screws 4 92 125 TOP VIEW PANEL INSTALLED 4 63 117 3 4 NPT 2 HLS 5 00 127 5 16 4 25 108 5 25 133 E M 5 00 127 To access ter...

Page 8: ...owmeter with DC Pulse Output Strip Char t Recorder 24 VDC Power Supply ISOLATED INPUT 4 20mA LOOP POWERED NOTE Separate power supplies are recommended to isolate output from input If isolation is not...

Page 9: ...the number of decimal locations allowable in the rate and total to acceptable ranges Press the key to move the decimal Press the E key to select the displayed decimal location NOTE For best performan...

Page 10: ...n GAL BBL MCF M3 blank Press the key to select the descrip tor Press the E key to enter the selected descriptor r sCale ratemeter scaling Sets the rate readout Choose rate per hour Hrs minutes nnIn or...

Page 11: ...divider may be needed so that the pulse output does not exceed the maximum rate of the pulse output If the pulse output pulses too quickly a ashing display will result Pressing the M key will result...

Page 12: ...y to step the factor decimal to the desired location Press the E key to enter the displayed decimal location Press the M key to skip and keep the existing location NOTE For best performance and resolu...

Page 13: ...s the value This will only display on units with Analog Output key to advance to the next digit Press the E key to enter the displayed value Press the M key to skip and keep the existing value Press t...

Page 14: ...ommon To reset the unit from the front panel the following key sequence is required Press M CLr tot will be displayed if the panel lock is on the display will prompt ent code Enter the proper code to...

Page 15: ...gital milliamp meter DMM in series with the analog output and simulating a full scale or over range ow rate If the output is out of calibration perform the following Remove power and or disconnect bat...

Page 16: ...r BATTERY POWERED Supplied with 1 or 2 C size Lithium battery pack EXTERNAL POWER INPUT Voltage 8 5 to 30 VDC Current Less than 5 mA Supplied with 1 C size lithium battery Protection Reverse Polarity...

Page 17: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 18: ...ty Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: