Chapter  2  Installation

2.1 Battery Installation:

All DPF500 models are shipped without the battery(ies) installed. This preserves battery life when the unit is not in 
service. When using external DPF5-BP, mount within 12" and plug connector into 3 position square posts (see Fig 
1). Polarity is not a concern because center is common.

To install the battery, begin by locating the battery holder. The DPF502 and DPF503 require opening the enclosure 
cover and removing the DPF500 to expose the battery holder.

The plus terminal of the battery is marked with a (+) symbol stamped into the battery holder. Be sure to install the 
battery(ies) correctly. 

Install battery(ies) to begin setup procedure. See Programming Flowchart to setup desired operating parameters.

2.2 Battery Replacement:

The DPF500 has a battery monitor feature which illuminates when the lithium battery voltage approaches its end 
of life. A descriptor, “BAT”, illuminates when the battery voltage falls below this predetermined value.  The low bat-
tery detector operates correctly with all power options. 

The battery, or batteries, should be replaced within several weeks of the fi rst occurrence of low battery warning, 
“BAT”.  Left unattended, the unit may become inaccurate, cease to operate or malfunction.

Before replacing the battery(ies), Press the 

 (left arrow) key to save the totalizer.  The display will show "

SAVE 

TOTAL

".  This will save the current total value and the total will resume from this value when the new battery(ies) 

is(are) installed.  

NOTE: 

 If the display starts to fl ash after the "

SAvE totAL

" message times out, press the "E" (enter) 

key.  If the message "E FLASH" is displayed, then there was not enough power left to save the setup and totalizer 
to fl ash memory.  At this point you must record the totalizer and setup  information and re-enter the setup data after 
the new battery(ies) is(are) installed.

Install new battery(ies) as described above.

Fig 1

6

5

4

3

2

1

Rear View

Battery

7

8

9

10

11

12

DPF5-BP

Connector

Summary of Contents for DPF500 Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info DPF500 Series Ratemeter Totalizer Shop online at User sGuide...

Page 2: ...78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056...

Page 3: ...nsions 3 2 4 Wiring 4 2 5 Typical Applications Wiring Diagrams 4 Chapter 3 20 Point Linearization Operation 3 1 20 Point Linearization LIN option 5 Chapter 4 Programming 4 1 Explanation of Display Pro...

Page 4: ...ii THIS PAGE INTENTIONALLY BLANK...

Page 5: ...Optically Isolated Displays Rate Total Simultaneously 5 Digit Rate Display 8 Digit Totalizer Display 4 20mA Analog Output MA option Powered From Internal Battery or 4 20 mA Output Loop MA option 20 P...

Page 6: ...its end of life A descriptor BAT illuminates when the battery voltage falls below this predetermined value The low bat tery detector operates correctly with all power options The battery or batteries...

Page 7: ...emove cover and 4 panel screws Terminals are on bottom side of PC board Panel Screws 4 92 125 TOP VIEW PANEL INSTALLED 4 63 117 3 4 NPT 2 HLS 5 00 127 5 16 4 25 108 5 25 133 E M 5 00 127 To access ter...

Page 8: ...owmeter with DC Pulse Output Strip Char t Recorder 24 VDC Power Supply ISOLATED INPUT 4 20mA LOOP POWERED NOTE Separate power supplies are recommended to isolate output from input If isolation is not...

Page 9: ...the number of decimal locations allowable in the rate and total to acceptable ranges Press the key to move the decimal Press the E key to select the displayed decimal location NOTE For best performan...

Page 10: ...n GAL BBL MCF M3 blank Press the key to select the descrip tor Press the E key to enter the selected descriptor r sCale ratemeter scaling Sets the rate readout Choose rate per hour Hrs minutes nnIn or...

Page 11: ...divider may be needed so that the pulse output does not exceed the maximum rate of the pulse output If the pulse output pulses too quickly a ashing display will result Pressing the M key will result...

Page 12: ...y to step the factor decimal to the desired location Press the E key to enter the displayed decimal location Press the M key to skip and keep the existing location NOTE For best performance and resolu...

Page 13: ...s the value This will only display on units with Analog Output key to advance to the next digit Press the E key to enter the displayed value Press the M key to skip and keep the existing value Press t...

Page 14: ...ommon To reset the unit from the front panel the following key sequence is required Press M CLr tot will be displayed if the panel lock is on the display will prompt ent code Enter the proper code to...

Page 15: ...gital milliamp meter DMM in series with the analog output and simulating a full scale or over range ow rate If the output is out of calibration perform the following Remove power and or disconnect bat...

Page 16: ...r BATTERY POWERED Supplied with 1 or 2 C size Lithium battery pack EXTERNAL POWER INPUT Voltage 8 5 to 30 VDC Current Less than 5 mA Supplied with 1 C size lithium battery Protection Reverse Polarity...

Page 17: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 18: ...ty Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: