Chapter  5  Operation

5.1  General

A suitable pulse producing device or fl ow meter is wired to one of the three pulse inputs provided on the DPF500. 
Only one of these inputs is used in a given application. There are no connections to the two unused pulse inputs.

Isolated magnetic pickups may be connected to terminals 1 and 2. Isolated contact closures may be connected to 
terminals 5 and 6. Any high level, DC 
pulse type may be connected to terminals 9(-) and 10(+).

Power to the unit may be provided by internal batteries, external DC, or the current loop. In all cases, the internal 
battery will provide for continued operation in the event primary power is lost.

Once properly wired, the operation of the DPF500 is automatic.

The fl ow totalizer is updated once per second* with battery power, instantaneously with DC or loop power.   If no 
input counts are received the unit remains in a low power state to conserve power.

The fl ow total may be cleared by the front panel switch sequence or by a contact closure on the remote reset ter-
minal to circuit common.

To reset the unit from the front panel, the following key sequence is required:

Press 

M   

“CLr tot” will be displayed   (if the panel lock is on, the display will prompt "

ent code

".  

 

Enter the proper code to advance to the 

clr tot

 prompt)

Press 

E   

To clear the total. Unit will return to operation

The fl ow rate indicator will measure the fl ow rate once every second* with battery power, 8 times per second with 
DC or loop power and display the fl ow rate. 

If the input pulses are not detected within the delay setting (0.1 to 99.9 seconds), a fl ow rate of 0 will be indicated.

The analog output will be scaled based on the user selected zero and full scale and the measured fl ow rate. The 
analog output is updated at the same time as the rate display.  

The pulse output updates at the same rate as the total display in accordance with the instrument setup of pulse 
scaling.

* A large delay setting and internal math operations may delay the update rate. A faster update rate occurs when the 
unit is loop powered or externally powered.

5.2 Maintenance

The only scheduled maintenance for the DPF500 is periodic replacement of the battery.

Battery Replacement:
See Section 2.1 for battery installation instructions.
See Section 2.2 for battery replacement instructions.

Summary of Contents for DPF500 Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info DPF500 Series Ratemeter Totalizer Shop online at User sGuide...

Page 2: ...78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056...

Page 3: ...nsions 3 2 4 Wiring 4 2 5 Typical Applications Wiring Diagrams 4 Chapter 3 20 Point Linearization Operation 3 1 20 Point Linearization LIN option 5 Chapter 4 Programming 4 1 Explanation of Display Pro...

Page 4: ...ii THIS PAGE INTENTIONALLY BLANK...

Page 5: ...Optically Isolated Displays Rate Total Simultaneously 5 Digit Rate Display 8 Digit Totalizer Display 4 20mA Analog Output MA option Powered From Internal Battery or 4 20 mA Output Loop MA option 20 P...

Page 6: ...its end of life A descriptor BAT illuminates when the battery voltage falls below this predetermined value The low bat tery detector operates correctly with all power options The battery or batteries...

Page 7: ...emove cover and 4 panel screws Terminals are on bottom side of PC board Panel Screws 4 92 125 TOP VIEW PANEL INSTALLED 4 63 117 3 4 NPT 2 HLS 5 00 127 5 16 4 25 108 5 25 133 E M 5 00 127 To access ter...

Page 8: ...owmeter with DC Pulse Output Strip Char t Recorder 24 VDC Power Supply ISOLATED INPUT 4 20mA LOOP POWERED NOTE Separate power supplies are recommended to isolate output from input If isolation is not...

Page 9: ...the number of decimal locations allowable in the rate and total to acceptable ranges Press the key to move the decimal Press the E key to select the displayed decimal location NOTE For best performan...

Page 10: ...n GAL BBL MCF M3 blank Press the key to select the descrip tor Press the E key to enter the selected descriptor r sCale ratemeter scaling Sets the rate readout Choose rate per hour Hrs minutes nnIn or...

Page 11: ...divider may be needed so that the pulse output does not exceed the maximum rate of the pulse output If the pulse output pulses too quickly a ashing display will result Pressing the M key will result...

Page 12: ...y to step the factor decimal to the desired location Press the E key to enter the displayed decimal location Press the M key to skip and keep the existing location NOTE For best performance and resolu...

Page 13: ...s the value This will only display on units with Analog Output key to advance to the next digit Press the E key to enter the displayed value Press the M key to skip and keep the existing value Press t...

Page 14: ...ommon To reset the unit from the front panel the following key sequence is required Press M CLr tot will be displayed if the panel lock is on the display will prompt ent code Enter the proper code to...

Page 15: ...gital milliamp meter DMM in series with the analog output and simulating a full scale or over range ow rate If the output is out of calibration perform the following Remove power and or disconnect bat...

Page 16: ...r BATTERY POWERED Supplied with 1 or 2 C size Lithium battery pack EXTERNAL POWER INPUT Voltage 8 5 to 30 VDC Current Less than 5 mA Supplied with 1 C size lithium battery Protection Reverse Polarity...

Page 17: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 18: ...ty Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: