3.2 Connection of Infrared and Blue Light Filter Unit
19
MAJ-2308 INSTRUCTION MANUAL
Ch.3
3.2
Connection of Infrared and Blue Light Filter Unit
1
Check the position of the brake pedal, push it
until it hits the back, apply the brake, and
confirm that the surgical microscope is locked
to the floor.
Figure 3.1
2
Confirm that the surgical microscope and LED light source are turned OFF by
referring to their instruction manual.
3
Using the provided Allen screwdriver (1.5 mm width across flats), loosen the lock
screws (× 2) of the cover of the microscope section completely.
Figure 3.2
CAUTION
• Always use the Allen screwdriver (1.5 mm width across flats) provided with this
instrument in the work. Otherwise, equipment damage may result.
• Hold the cover securely with a hand during the work. Otherwise, the cover may
drop and get damaged.
Brake pedal
Allen screwdriver
(1.5 mm width across flats)
Lock screws
The cover of the
microscope
section