Summary of Equipment Functions
MAJ-2301 INSTRUCTION MANUAL
This instrument is designed to be connected to the Olympus-designated surgical microscope. By
pairing between the foot switch and wireless unit, the control signals of the surgical microscope can be
transmitted with either wireless or wired connection. The function incorporated in this instrument is
outlined below.
Note that only one foot switch can be pairing with a wireless unit.
Pairing of foot switch and wireless unit
Pairing of the foot switch and wireless unit enables wireless communication between them.
Pairing can be set using the foot switch cable provided with this instrument. Wireless control of
the functions of the surgical microscope is possible by controlling the foot switch.
Refer to Section 3.6, “Pairing”.
Wired connection of foot switch
In case the wireless signal is weak or the batteries are exhausted, the foot switch can be
connected directly to the Olympus-designated surgical microscope using the provided foot
switch cable. When the foot switch is connected through the foot switch cable, the functions of
the surgical microscope can be controlled from the foot switch even when it contains no
batteries in it.
Refer to Section 5.3, “Wired control of foot switch”.
NOTE
• The foot switch cannot control the microscope functions unless the provided
wireless unit is connected to the microscope. Be sure to connect the wireless unit
to the surgical microscope even before the wired control operation.
• Wireless communication is not performed when wired connection.