3.6 Pairing
29
MAJ-2301 INSTRUCTION MANUAL
Ch.3
CAUTION
• Perform pairing before using foot switch. Without pairing, the foot switch becomes
incapable of wireless control of the surgical microscope or there may even be a
case in which a function of an unintended surgical microscope is activated.
• Be sure to turn off the surgical microscope before connecting the foot switch cable.
Otherwise, equipment failure or malfunction may result.
NOTE
In the battery exchange after pairing, re-pairing is not necessary.
3.6
Pairing
1
Confirm the power of the surgical microscope is off.
2
Open the terminal cover on the foot switch, hold the FOOTSWITCH connector (on the
side marked “FOOTSWITCH”) of the foot switch cable so that the index faces upward
and insert the connector into the terminal on the foot switch until it clicks.
Figure 3.14
Foot switch marking
Index
Terminal
cover
Connection
terminal
Foot switch connector
Foot switch
Foot switch cable